Trace Oxygen Analyzer
The connectors are:
Threshold Alarm 1:
Can be configured as high (actuates when concentration is above
threshold), or low (actuates when concentration is below
threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
Threshold Alarm 2:
Can be configured as high (actuates when concentration is above
threshold), or low (actuates when concentration is below
threshold).
Can be configured as failsafe or non-failsafe.
Can be configured as latching or non-latching.
Can be configured out (defeated).
System Alarm:
Actuates when DC power supplied to circuits is unacceptable in
one or more parameters.
Permanently configured as failsafe and latching. Cannot be
defeated.
Actuates if self test fails.
Teledyne Analytical Instruments
Figure 3-6: Types of Relay Contacts
Installation
25