Figure 1-3: Model 3020T Gas Connector Panel - Teledyne 3020T Operating Instructions Manual

Trace oxygen analyzer
Hide thumbs Also See for 3020T:
Table of Contents

Advertisement

Introduction

Figure 1-3: Model 3020T Gas Connector Panel

Gas Inlet and Outlet One inlet (must be externally valved)
Optional:
Calibration Gas Ports Separate fittings for zero, span and
Note: If you require highly accurate Auto-Cal timing, use external
Auto-Cal control where possible. The internal clock in the
Model 3020T is accurate to 2-3%. Accordingly, internally
scheduled calibrations can vary 2-3% per day.
Teledyne Analytical Instruments
and one exhaust out.
sample gas input, plus internal
valves for automatically switching
the gases in sync with the 3020
electronics.
Model 3020T
8