Maintenance of a Pump Head with Seal Wash Option

When
When maintaining seal wash option
Tools required
• 4-mm hexagonal key
Parts required
#
1
1
1
Preparations
• Switch off pump at the main power switch
• Remove the front cover
• Use a solvent shutoff valve or lift up solvent filters for avoiding leakages
• Remove the pump head, see
• Remove the wash solvent tubings from the support ring inlet and outlet
Agilent 1260 IsoPump/QuatPump User Manual
p/n
Description
01018-23702
Insert tool
0905-1175
Wash seal
5062-2484
Gasket, seal wash (pack of 6)

Maintenance of a Pump Head with Seal Wash Option

"Removing the Pump Head Assembly"
Maintenance
on page 128