Maintenance Of A Pump Head Without Seal Wash Option - Agilent Technologies 1260 Infinity User Manual

Isocratic pump and quaternary pump
Hide thumbs Also See for 1260 Infinity:
Table of Contents

Advertisement

9
Maintenance

Maintenance of a Pump Head Without Seal Wash Option

Maintenance of a Pump Head Without Seal Wash Option

When
In case of maintenance or pump head internal leaks.
Tools required
• Wrench 1/4 inch
• 4-mm hexagonal key
Parts required
#
1
1
1
1
1
Preparations
• Switch off pump at the main power switch
• Remove the front cover
130
p/n
Description
01018-23702
Insert tool
5063-6589
Standard seals (pack of 2)
or
0905-1420
PE seals (pack of 2)
5063-6586
Piston
"Removing the Pump Head Assembly"
on page 128
Agilent 1260 IsoPump/QuatPump User Manual