Performance Tests; Required Equipment List - Fluke 525A Service Manual

Temperature/pressure calibrator
Hide thumbs Also See for 525A:
Table of Contents

Advertisement

Performance Tests

The following performance tests are used to verify the functionality of the Calibrator. If
the Calibrator fails any part of the performance tests, calibration adjustment or repair by a
Fluke Service Center is in order. For a list of service centers, refer to "Contacting Fluke"
earlier in this manual.

Required Equipment List

To complete the performance tests and calibration adjustment, the equipment listed in
Table 10 is necessary. If the equipment listed is not available, equipment with the same
or better specifications can be substituted.
Manufacturer
Fluke
Fluke- 525A/Leads
Fluke
742A-10
Fluke
742A-100
Fluke
742A-10K
Fluke
5520A
Wavetek
1281
_
_
_
_
Hart Scientific
1521
Omega
_
_
_
_
_
_
Table 10. Required Equipment
Model
Equipment
Test Lead Set
Resistance
Standard, 10 Ω
Resistance
Standard, 100 Ω
Resistance
Standard, 10K Ω
Calibrator
DMM
Ohms Shorting
Block
RTD Probe
Precision
Thermometer
J Thermocouple
Probe
Lag Bath
Dewar Flask with
lid
Banana Jack to
Copper TC Mini-
Connector Cable
Purpose
All Functions
100 mA source calibration
Low Ohms Measure and SPRT
Calibration.
Hi Ohms Measure Calibration
Low Ohms Measure and SPRT
Calibration. Hi Ohms Measure
Calibration.
All functions except CJC
Low Ohms Measure and SPRT
Calibration. Hi Ohms Measure
Calibration.
TC CJC Calibration
TC CJC Calibration
TC CJC Calibration
TC CJC Calibration
TC tests
TC tests
525A
Performance Tests
23