Service Information

Each Calibrator is warranted to the original purchaser for a period of one year beginning
on the date received. The warranty is located at the front of this manual.
Factory-authorized service and technical advice for the Calibrator is available at Fluke
Service Centers. A complete list of service centers is available at www.fluke.com.
Servicing described in this manual is to be done by qualified
service personnel only. To avoid electrical shock, do not
service the Calibrator unless you are qualified to do so.

Placement and Rack Mounting

Place the Calibrator on top of a bench or mounted in a standard-width, 24-inch (61-cm)
deep equipment rack. For bench-top use, the Calibrator is equipped with non-slipping,
non-marring feet. To mount the Calibrator in an equipment rack, use the Rack Mount
Kit, Model Y5320A, instructions are included with the kit. For convenience, the rack
mount instruction sheet can be stored in the binder of this manual.

Cooling Considerations

Damage caused by overheating may occur if the area around
the air intake is restricted, the intake air is too warm, or the fan
filter becomes clogged.
Accuracy and dependability of all internal parts of the Calibrator are enhanced by
maintaining the coolest possible internal temperature. By observing the following rules,
you can lengthen the life of the Calibrator:
• The area around the fan filter must be at least 3 inches from nearby walls or rack
enclosures.
• Exhaust perforations on the sides of the Calibrator must be clear of obstructions. Most
of the heat exits from the side vents.
• Air entering the instrument must be room temperature. Make sure that exhaust from
another instrument is not directed into the fan inlet.
Clean the fan filter every 30 days, or more frequently if the Calibrator is operated in a
dusty environment. (Instructions for cleaning the fan filter are in Chapter 2 of the
Users Manual.)

Selecting Line Voltage

The Calibrator will operate on two different input line voltages. The present line voltage
setting is indicated on the outside of the line voltage switch found on the Calibrator's
rear panel.
To change the line voltage:
1. Disconnect the Calibrator from line power by unplugging the power cable.
2. Using a flat-blade screwdriver, rotate the switch until the desired voltage is under
3. Attach the Calibrator to the power source using a power cable that will mate with
12
the arrowhead on the line voltage switch.
power source outlet.
XWWarning
WCaution