High and Low Blood Sugar Alert Screens
Your meter has been preset with a low blood sugar (hypoglycaemia) value of 3.9 mmol/L
and a high blood sugar (hyperglycaemia) value of 13.9 mmol/L. These are the preset values,
but can be customized by you and/or your health care professional. You can change the
High and Low Alert levels under Main Menu, Setup option (see page 51).
If your blood sugar reading is under your low blood sugar alert level:
3.8
Low Blood Sugar
Low Blood Sugar
If your blood sugar reading is over your high blood sugar alert level:
14.0
High Blood Sugar
10:30 | 13.10
Bolus
Bolus
14.0
mmol
Reminder
Reminder
L
High Blood Sugar
Notes
Notes
ADD NOTES
Before Meal
OK
After Meal
Time After Meal
Don't Feel Right
Sick
Stress
Activity
Go Back
Testing in the Dark
Your meter has a lighted test strip port to help you test in the dark.
• With the meter off, give the top Menu button two quick presses to turn on the test strip
port light.
• Insert a test strip and the display screen appears.
• Once blood is applied to the test strip, the light goes off.
• Continue with your test.
• Two quick presses of the top Menu button also turns off the light.
®
22
84647013_CntrNextLINK2_4_UG_EN_FpBp_v3.indd 22-23
84647013_CntrNextLINK2_4_UG_EN_FpBp_v3.indd 22-23
• A screen with large orange numbers alerts you that
your blood sugar is low.
• A screen with large orange numbers alerts you that
your blood sugar is high.
• If AutoLog is ON and you did not mark your result
before the high or low alert appeared, select Notes.
• Press the L or M button to scroll through the choices.
• Press OK when your choice is highlighted.
Control Solution* Testing
WARNING
Shake the control solution well before testing.
CAUTION: Use only C
®
N
control solution from Bayer with your
ONTOUR
EXT
C
®
N
LINK 2.4 blood glucose monitoring system. Using anything other than
ONTOUR
EXT
C
N
control solution can cause inaccurate results.
ONTOUR
EXT
* Control solution may not be included in all meter kits. Contact Bayer Diabetes Support for information
on obtaining control solution. See back cover for contact information.
Quality Control
You should perform a control test when:
• using your meter for the first time
• you open a new bottle or package of test strips
• you think your meter may not be working correctly
• you have repeated, unexpected blood glucose results
WARNING
• Do not calibrate your continuous glucose monitoring device from a control result
• Do not calculate a bolus based on a control result
CAUTION: Check the expiry date on the test strip bottle and the expiry date and
discard date on the control solution bottle. DO NOT use expired materials.
Control solutions are available separately if not included in the meter kit. Always use
C
N
control solutions from Bayer. Other brands could present incorrect results. If
ONTOUR
EXT
you need help locating C
N
control solutions from Bayer, contact Bayer Diabetes
ONTOUR
EXT
Support. Please see back cover for contact information.
1. Remove the test strip from the bottle and firmly snap the lid closed.
2. Hold the test strip with the grey square end facing up.
3. Insert the grey square end of the test strip into the test
strip port until the meter beeps.
APPLY BLOOD
4. You will see the Apply Blood screen. The meter is now
ready for you to test.