Technical Information - Bayer HealthCare Contour Next Link 2.4 User Manual

Wireless blood glucose monitoring system
Hide thumbs Also See for Contour Next Link 2.4:
Table of Contents

Advertisement

Symptoms of High or Low Blood Sugar
You can better understand your test results by being aware of the symptoms of high or low
blood sugar. Some of the most common symptoms are:
Low blood sugar
(Hypoglycaemia):
• Shakiness
• Frequent Urination
• Sweating
• Excessive Thirst
• Fast Heartbeat
• Blurred Vision
• Blurred Vision
• Increased Fatigue
• Confusion
• Hunger
• Passing Out
• Irritability
• Seizure
• Extreme Hunger
• Dizziness
If you are experiencing any of these symptoms, test your blood sugar. If your
test result is under 2.8 mmol/L or above 13.9 mmol/L, contact your health care
professional immediately.
For additional information and a complete list of symptoms, contact your health
care professional.
®
56
84647013_CntrNextLINK2_4_UG_EN_FpBp_v3.indd 56-57
84647013_CntrNextLINK2_4_UG_EN_FpBp_v3.indd 56-57
High blood sugar
Ketones (Ketoacidosis):
(Hyperglycaemia):
• Shortness of Breath
• Nausea or Vomiting
• Very Dry Mouth
WARNING

Technical Information

Accuracy
The C
®
N
LINK 2.4 blood glucose monitoring system was tested in 100 capillary
ONTOUR
EXT
blood samples using 600 C
®
N
test strips. Two replicates were tested with each
ONTOUR
EXT
of 3 lots of C
N
test strips for a total of 600 readings. Results were compared to
ONTOUR
EXT
the YSI
®
glucose analyzer, which is traceable to the CDC hexokinase method.
below compare the performance of the two methods.
Table 1 — System accuracy results for C
ONTOUR
C
N
test strips at glucose concentration < 4.2 mmol/L
ONTOUR
EXT
Within +/–
Within +/–
0.28 mmol/L
0.56 mmol/L
65 / 78
75 / 78
(83.3%)
(96.2%)
Table 2 — System accuracy results for C
ONTOUR
C
N
test strips at glucose concentration ≥ 4.2 mmol/L
ONTOUR
EXT
Within +/– 5%
Within +/– 10%
396 / 522
517 / 522
(75.9%)
(99.0%)
Acceptance criteria in ISO 15197: 2003 are that 95% of all differences in glucose values
(i.e., between reference method and meter) should be within +/-0.83 mmol/L for glucose
values less than 4.2 mmol/L, and within +/- 20% for glucose values greater than or equal to
4.2 mmol/L.
Precision
A repeatability study was conducted with the C
ONTOUR
monitoring system using 5 venous whole blood specimens with glucose levels from
2.2 to 18.5 mmol/L. Multiple replicates (n=100 per lot) were tested using multiple
C
N
 LINK 2.4 blood glucose meters and 3 lots of C
ONTOUR
EXT
following precision results were obtained.
System repeatability results for C
N
ONTOUR
EXT
test strips
Mean Glucose Level,
Standard Deviation,
mmol/L
mmol/L
2.20
0.04
4.48
0.06
6.79
0.09
11.37
0.16
18.34
0.25
8
The tables
N
LINK 2.4 meter using
EXT
Within +/–
0.83 mmol/L
78 / 78
(100%)
N
LINK 2.4 meter using
EXT
Within +/– 15%
Within +/– 20%
522 / 522
522 / 522
(100%)
(100%)
N
EXT
LINK 2.4 blood glucose
N
test strips. The
ONTOUR
EXT
LINK 2.4 meter using C
N
ONTOUR
EXT
Coefficient of Variation
(%)
2.0
1.3
1.3
1.4
1.4
57
3/18/13 8:05 AM
3/18/13 8:05 AM