local Bayer Diabetes Care Customer Service.
WHAT YOU SEE
Some display seg-
ments do not appear
when the meter
initially turns on.
Control test result
is out-of-range (too
high or too low).
WHAT IT MEANS
Possible meter
electronics failure.
Test strip is deterio-
rated due to expo-
sure to humidity or
heat.
Control solution
is past expiration
date or is past the
open use date.
Test strip is past
expiration date or
is past the open
use date.
Incorrect control
solution was used.
Control solution
is not at room
temperature.
Control solution is
not mixed well.
www.bayerdiabetes.com
WHAT YOU SHOULD DO
Press and hold
while the
meter is off. Compare meter dis-
play with the picture on page 3.
If the display is still not correct,
call Customer Service.
Run another control test with
a new test strip.
Check all expiration dates.
Do not use expired testing
materials.
Make sure you are using
C
TS control solution.
ONTOUR
Wait 20 minutes until control
solution is at room tempera-
ture. Try again.
Gently rock the control bottle
to ensure the control solution
is mixed well.
If result is still out-of-range,
retest with a new test strip
and control solution.
If result is still out-of-range,
call Customer Service.