Diagnostics

Problem description:
Cause:
Remedy:
Problem description:
Cause:
Remedy:
Problem description:
Cause:
Remedy:
The SpectroScan T does not work, and power supply
indicator lights do not come on.
No electrical supply
Check that the cable and connections are correctly
inserted, and that the device is switched on. Replace the
fuses (see chap. 5.4.3.3.).
Communication with hardware impossible.
Modified baud rate.
Check configuration of the DIP-switches and correct cable
connections.
Perform a reset as described in chapter 5.4.3.1.
The transverse and longitudinal slides do not move, or
make strange noises when they move.
Software or motor problem.
Remove the transportation protection (chap. 2.3.2)
Perform a reset as described in chapter 5.4.3.1., then
perform a demo (chap. 5.4.3.2).