Diagnostics - GMB Spectrolino Operation Manual

Table of Contents

Advertisement

3.5

Diagnostics

Problem description:
Cause:
Remedy:
Problem description:
Cause:
Remedy:
Problem description:
Cause:
Remedy:
The Spectrolino does not work.
No electrical supply.
Check to see that the charging unit is correctly inserted
into the GretagMacbeth cable or the mini-DIN adapter.
Communication with hardware impossible.
Modified Baud rate.
Check that cables and terminals are correctly inserted.
Perform a reset as indicated in chap. 3.4.3
Measured values are unrealistic
Measurement conditions varied, or the measuring lamp
is malfunctioning
Check that the correct tube is being used, as well as
the white reference belonging to the measuring device.
Check that the measuring lamp lights up by performing a
simple measurement.
Perform a reset as indicated in chap. 3.4.3
Spectrolino
3
operation
19
EN