5.4.3.4 Replacing the measurement lamp
To replace the measurement lamp, turn off SpectroScan T and remove the aperture,
as described on chapter 5.4.2.3. Pull down the lamp to remove it from the receptacle.
Extract the plug that connects the lamp to the instrument. Set the new lamp so that
the cable and the guidance pin are turned to the right, then carefully insert the new
lamp. Ensure that the guidance pin (see the arrow) is correctly set inside the groove.
Now connect the plug of the new lamp to the instrument and re-insert the measure-
ment aperture. When the instrument is turned on, it automatically performs a
re-positioning. Perform a calibration before starting a new measurement.
- Use exclusively GretagMacbeth lamps, article 36.61.16.
SpectroScan T
5
operation
41
EN