SpectroScan
4
operation
26
EN
4.4
Use
The SpectroScan must always be operated with the appropriate software. This
document explains procedures for applying the software supplied with the device,
called SpectroChart Lite. Specific information regarding this software is given in the
"Readme.txt" file in the program.
4.4.1 Measurement original setup
Position the test chart to read on the electrostatic surface (37), such that it does not
exceed the white markings. Pressing the field command key (36)
test chart to the measuring table.
- The allowed thickness for test originals is just 1.5 mm. To avoid measuring
errors, this value must not be exceeded.
- Normally no supplementary fixing is required for the original.
- Ensure that the original is level, and that no bubbles form.
- The orientation of the original is irrelevant, as any obliquity is
compensated by the software.
4.4.2 Definition of original type and orientation
Initiate the SpectroChart Lite software by double-clicking on the corresponding icon.
The software itself establishes the link with the SpectroScan and checks that the
Spectrolino is connected. If communication is correctly established, it is necessary to
inform the software which type of test chart is being used. Clicking on the field
displays a selection of the available test charts. Select the desired test chart and
confirm selection by clicking
O.K
.
, fixes the
Pap.hold
measure