11.3 Weighing units switch-over

The weighing unit selected here remains even after disconnection from the mains.
Press UNIT repeatedly until the desired setting is displayed.
To enable or disable individual units go to the menu (See chap. 9.1 menu overview)
For setting options see chap. 12.7
Weighing scales with type approval options are restricted to [g]
[ct].

11.4 Taring

The dead weight of any weighing container may be tared away by pressing a button,
so that the following weighing procedures show the net weight of the goods to be
weighed.
Put vessel of goods to be weighed on weighing plate and close the wind
screen doors.
Wait until the stability display appears (
the container is now internally saved.
Weigh the goods to be weighed and close the wind screen doors.
Wait until the stability display appears (
Read net weight.
When the balance is unloaded the saved taring value is displayed with
negative sign.
To delete the stored tare value, remove load from weighing plate and
press TARE.
The taring process can be repeated any number of times. The limit is
reached when the whole weighing range is exhausted.

11.4.1 Entering a tare value manually (pre-tare)

In weighing mode press F button
Press  to select setting ENTER TARE and confirm by pressing PRINT.
Use the arrow keys for entering the tare value and confirm by pressing PRINT.
The entered tare value will be displayed with a "

11.4.2 Deleting the tare value

To delete the tare value, press the TARE-key or enter value"0.000".
36
), then press TARE. The weight of
).
-
"
ILJ-C/ILS-C-BA-e-1320
[kg]