The information strip shows the two limits. The MIN symbol will appear at the upper
margin of the display as 0 is the point from which values drop below the lower limit.
The weighing scale is ready for control weighing. Use the symbols MIN-OK-MAX to
check whether the load is within, above or below the intended tolerance.
If invalid values are entered such as lower tolerance limit greater than
upper tolerance limit, the balance will issue the error message and
return automatically to weighing mode.
Return to weighing mode
Press MODE and F1 WEIGHING will be displayed. Confirm by pressing PRINT and
the weighing scale will change to weighing mode.

13.3 Dosing

If dosing mode is enabled, the weighing scale will work at increased display speed.
Please keep in mind that the weighing scale reacts very sensitively to ambient
conditions.
Enter the target weight (reference weight) along with the tolerance:
The ENTER REFERENCE WEIGHT function is allocated to the F4-key.
ILJ-C/ILS-C-BA-e-1320
Both limits have now been entered;
the display changes to control
weighing mode.
Select dosing mode
4x