13.4 Percent determination

Percent determination allows weight display in percent, in relation to a reference
weight. The reference weight may be determined by weighing or by entering it
numerically.
Determining reference by weighing
The DETERMINE REFERNCE WEIGHT function is allocated to the F1-key.
Go to menu and select P2.6 DEVIATION
Use the arrow keys to select P2.6.5 SHORTCUTS
Press F2, P2.6.5.1 will be displayed along with the previous function for the
F1-key.
Again press F2 and the information line will start flashing.
Use the arrow keys to select DETERMINE REFERENCE WEIGHT and
Confirm by PRINT; this function has now been allocated to F1.
In weighing mode:
(example)
Remove reference weight and place load. The weight will be displayed in
percent according to reference weight.
62
This completes setting the operating
mode weighing by percentage.
The percent symbol will be
displayed at the upper margin of the
display. The setting REFERENCE
WEIGHT 0.0 G will be displayed at
the lower margin.
Put the reference weight equalling
100% on the weighing plate.
Example: 1 kg is supposed to be
equivalent to 100%.
The reference weight of 1 kg has now been saved
as 100% value.
ILJ-C/ILS-C-BA-e-1320