7.2 Menu settings

Menu item „P1 ADJUSTMENT"
ILJ-C
P1.1
INT.JUST.
-
-
P1.4
TEST JUST.
* = Locked in weighing scales with type approval certificate [Er 9 lock].
7.2.1 P1.1 INT.JUST. – Internal adjustment (models ILJ-C only)
For automatic internal adjustment see chap. 7.1
Starting internal adjustment manually
Press the CAL-key or invoke menu "P1.1 INT.JUST."
Press
or
Ensure that there are no objects on the weighing plate.
You can cancel the adjustment by pressing the UNIT-key.
After successful adjustment the weighing scale will return automatically to
the menu item. To return to weighing mode, press UNIT.
ILJ-C/ILS-C-BA-e-1320
[Adjustment]
ILS-C
-
P1.2
EXT.JUST.
P1.3
BE.JUST.
-
Internal adjustment will be taking place.
Internal adjustment, ILJ-C only
External adjustment *
User adjustment
Adjustment test