MEANING: The test has been performed outside the operating temperature range. ACTIONS:
Move to an area with an ambient temperature of 10°C to 40°C. Wait for the meter and test strips
to reach the new temperature (usually 10-20 minutes) and retest.
LOW BATTERY MESSAGE: Flashing message
MEANING: Battery #2 in your meter is low.
ACTIONS: If the low battery symbol appears, battery #2 must be replaced. Battery numbers are
indicated by the numbers printed on the pull tabs.
DEAD BATTERY MESSAGE: Flashing message
MEANING: Battery #2 in your meter is dead.
ACTIONS: If the dead battery symbol appears, the dead battery must be replaced.
LOW BACKLIGHT BATTERY: Backlight flashes when user presses the BACKLIGHT (
)
button.
MEANING: Battery #1, which powers the backlight, is low.
ACTIONS: If the backlight flashes when user presses the BACKLIGHT (
) button, battery #1
must be replaced. If the backlight does not turn on, battery #1 is dead and must be replaced.
(Battery numbers are indicated by the numbers printed on the pull tabs).

21 Batteries

CHAPTER
CAUTION: When the battery symbol appears, you should replace the batteries immediately.
Only use CR2032 batteries.
Your BGStar Meter comes with two pre-installed, CR2032, 3 volt, lithium batteries. Battery #1
powers the backlight and battery #2 powers the basic meter functions. When the power from
battery #2 is low, the low battery symbol or dead battery symbol will be displayed on the screen
until you change battery #2. When replacing battery #2, the time and date information may be lost
after the battery is removed from the meter. If this happens, activate the meter after installing the
new battery and it will prompt you to set the time and date. Correct time and date settings are
needed to get correct averages. Removing the batteries does not affect the meter's memory or
previously stored user settings, such as time format, alarms, etc.
Note: Dispose of used batteries according to your local environmental regulations. Your meter is
an electronic device. When disposing of your meter follow all local environmental regulations.