Bring the tip of the test strip to the control solution sample. The BGStar System will
automatically detect if a test is performed with BGStar Control Solution.
When control solution is applied, the visual fill window on the BGStar Test Strip turns blue. Your
meter will beep once indicating that you should remove the tip of the BGStar Test Strip from the
control solution sample. On the meter display, the droplet symbol fills. The scrolling bar and 1-2-
3 indicates that the meter is calculating your results.
Droplet
:
Symbol
[4] Viewing Control Solution Test Result: The result will appear on the display and, if the
volume is enabled, your meter will beep. The date and the abbreviation "Ctrl" will also appear on
the display.
Compare the result of your control solution test to the range printed on the BGStar Test Strip vial
label for the control solution level that you used. Your control solution result should fall within
this range. The control solution test result is stored in memory.
Application number: 184338
Date of revision: August 18, 2011
s-a Version 2.1 dated August 26, 2011
This is the visual fill window.
DO NOT PUT CONTROL
SOLUTION HERE.
Apply control solution here.