Normal: 4.4-6.7 mmol/L
FOR EXAMPLE ONLY.
Refer to your test strip vial
label for corresponding
range.
To deactivate the meter, remove the used BGStar Test Strip. Dispose of used materials properly.
If your control solution test results are out of the range, repeat the test. If the results from the
control solution tests continue to fall outside that range, do not use the BGStar System to test your
blood glucose. Call sanofi-aventis toll free at 1-888-852-6887.
IMPORTANT: Out-of-range test results may be due to one or more of the following factors:
[1] Expired or defective control solution.
[2] Expired or defective test strip.
[3] Error in performing test.
[4] Watered-down control solution.
[5] Meter malfunction.
[6] Control solution test done outside the system operating temperature of 10°C to 40°C.
[7] Failure to shake the control solution bottle vigorously before using.
[8] Failure to discard the first drop of control solution and to wipe the bottle tip clean.
CAUTION: Results from control solution tests do not reflect your blood glucose level. The
control solution range is a target range for control solution only. It is not a target range for
your blood glucose level.
Control results are not included in the
pre and post meal time,
14, 30
and 90 day averages.
Application number: 184338
Page 15 of 48
Date of revision: August 18, 2011
s-a Version 2.1 dated August 26, 2011