Troubleshooting
PROBLEM
LED lamp does not light
The brightness of the LED lamp is
not adequate
®
ACCU-SCOPE
No power supply
The LED wire is disconnected
The LED lamp is damaged
The rechargeable battery is aging
The charging circuit is not working
73 Mall Drive, Commack, NY 11725 • 631-864-1000 • www.accu-scope.com
3002 MICROSCOPE SERIES
CAUSE
SOLUTION
Check the power cord
or recharge the battery
Reconnect the LED wire
OR
call your local authorized
ACCU-SCOPE dealer for service
Change the LED lamp assembly –
CAT# 02-3259
OR
call your local authorized
ACCU-SCOPE dealer for service
Replace the battery pack –
CAT# 02-1060
OR
call your local authorized
ACCU-SCOPE dealer for service
Replace the charging circuit board
OR
call your local authorized
ACCU-SCOPE dealer for service