7. Set Overload Stops as shown in Section 3.6.
3.6

Setting the Overload Stops

The Overload Stop gaps must be checked and reset if the Load Cell is replaced. This
procedure requires test masses equal to the scale's capacity. (See Table 3-2.)
There are four Overload Stop Bolts. (See Figure 3-5.)
Adjust the Overload Stops, per Table 3-2:
1. Adjust the Overload Stop Bolt so that
the gap between the Load Cell Frame
and the Overload Stop Nut is equal to
the specification in Table 3-2. (The
gap can be tested by applying 100%
load, one corner at a time, at each
corner stop. If the gap is right, the Load
Cell Frame will just touch the Overload
Stop Nut.)
Note: Be careful not to overload the
Load Cell, which would damage it.
2. Repeat this for all four corners.
3. Test the scale to see if full capacity
can be achieved.
Valor™ 2000/4000 Series Service Manual
TABLE 3-1. MOUNTING BOLT TORQUE SETTINGS
MODEL
V41xWE1501T
V22xWE1501T
V41xWE3T
V22xWE3T
V41xWE6T
V22xWE6T
V41xWE15T
V22xWE15T
Overload
Stops
CHAPTER 3 MAINTENANCE PROCEDURES
TORQUE SETTING
6N to 8N
6N to 8N
6N to 8N
6N to 8N
Figure 3-5. Overload Stops.
3-5
Ohaus Corporation www.ohaus.com