CHAPTER 3 MAINTENANCE PROCEDURES

3.3.1 Separating the Top and Bottom Housings

Common hand tools are sufficient to disassemble the Valor 2000/4000 scales. Turn the scale
off and unplug the power cord before you begin.
1. Turn the scale over.
2. Remove the 4 screws of battery cover.
3. Take away the battery cover.
4. Disconnect the battery.
5. Remove the battery.
6. Remove the 2 screws of pan and take away the pan.
7. Remove the LFT cover screws.
8. Take away the LFT cover.
9. Remove the 5 screws.
10. Turn the scale over and move the top housing a little bit
backward and then raise and remove
the top housing.
11. Separate Top Housing from Bottom
Housing. Avoid straining the cables
that connect the Main PCB to the
parts in the Bottom Housing. (Lay
the two housings close to each
other, so cable is not strained.)
Ohaus Corporation www.ohaus.com
Figure 3-1. Screws that secure the
housing.
3-2