Appendix A Standard Calibration; Calibration - OHAUS VALOR 4000 Service Manual

Table of Contents

Advertisement

APPENDIX A. STANDARD CALIBRATION & SETUP
A.1 CALIBRATION
Standard calibration should be performed prior to using a scale, and after service.
CAUTION:
Be careful not to touch the scale or the table while calibration is in progress,
as it will cause the process to fail.
Preliminary Steps:
1. Be sure the scale is level and stable during the entire calibration process.
2. Allow the scale to warm up for approximately five minutes after stabilizing to room
temperature.
3. To abort calibration, press Exit key or turn the scale off anytime during the calibration
process.
4. Before performing the calibration, be sure to have the appropriate calibration weights as
listed in table A-1(Valor 4000) and table A-2(Valor 2000).
5. Ensure that the LFT PINs lock is set to unlocked positions.
6. Or adjust the GEO setting according to your location (see instruction manual).
Valor™ 2000/4000 Series Service Manual
Table A-1. Span Calibration Mass for Valor 4000
Span Calibration Mass (sold separately)
Max
1500g
3000g
6000g
15000g
Table A-2. Span Calibration Mass for Valor 2000
Span Calibration Mass (sold separately)
Max
Mass
1500 g
1 kg
3000 g
2 kg
A-1

APPENDIX A STANDARD CALIBRATION

1
Mass
1.5kg / 3lb
3kg / 6lb
6kg / 15lb
15kg / 30lb
Max
Mass
6000 g
5 kg
15000 g
10 kg
Ohaus Corporation www.ohaus.com