Adjusting Off Center Load - OHAUS VALOR 4000 Service Manual

Table of Contents

Advertisement

CHAPTER 4 TESTING
5. Maximum allowable change in displayed weight readings for each of the four
positions can be found in Tables 1-1 (Specifications, page 1-2).
If this maximum is exceeded, follow procedures in Section 4.4.5, Adjusting Off
Center Load.

4.4.5 Adjusting Off Center Load

If the Off Center Load (OCL) is excessive, perform adjustment as follows:
B
Weighing Pan's center,
with points A – D indicated
Figure 4-1. Scale drawing of Valor 4000 Load Cell and Weighing Pan.
1. Place the test weight in the center of the Weighing Pan.
2. Tare the balance.
3. Move the weight to point A and record the reading.
4. Move the weight to point B and record the reading.
5. Move the weight to point C and record the reading.
6. Move the weight to point D and record the reading.
7. If the reading at point A is negative, file at points 1 and 4 AT AN ANGLE.
8. If the reading at point B is negative, file at points 1 and 2 STRAIGHT ACROSS.
9. If the reading at point C is negative, file at points 2 and 3 AT AN ANGLE.
10. If the reading at point D is negative, file at points 3 and 4 STRAIGHT ACROSS.
Note: It is not recommended that you try to adjust more than –5 counts if the beam has
been filed already. If the beam has not been filed previously, you can adjust –10 counts.
Remember, when filing you are weakening the beam. File a little at a time.
Ohaus Corporation www.ohaus.com
C
D
A
4-6
2
1
Top view of Load Cell
Pan
Side view of Load Cell
2
1
Valor™ 4000 Series Service Manual
3
4
Fixed end
3
4

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Valor 2000

Table of Contents