Calibration - A&D SB-30 Instruction Manual

Digital bench scales
Table of Contents

Advertisement

Setup Menu Mode /
NAME/CODE
DESCRIPTION
Selects display mode for the scale.
F1
Display Mode
Selects the range within which the scale may be zeroed. Note that
F6
the scale must be in standstill to zero the scale.
Zero Range
F9
Selects the primary base unit to be used in the calibration process.
Also the default unit for normal operation.
Calib. Unit
"0" = primary unit is lb.
Actuates the function which illuminates all digit segments, decimal
F11
points, and LCD annunciators in a test sequence. Pressing the
Display Check
ZERO key to scroll down one level begins the test sequence.
Places scale into the calibration routine. Scrolling down with the
F12
ZERO key one level begins the procedure.

Calibration

This sub-menu will reset all parameters in the "F" and "A" menus to
F13
the default settings. USE WITH CAUTION!!!!!
Factory Reset
Calibration
ON/OFF
DC POWER
SWITCH
JACK
ID TAG
CALIBRATION
SWITCH COVER
The scale can be calibrated by following the procedure shown below. The minimum test weight
and the recommended test weight depends on the scale's capacity.
For the SB-30, the minimum test weight is 0.3 lb (0.15 kg) , and the recommended test weight is
20 lb (10 kg). For the SB-70, the minimum test weight is 0.7 lb (0.35 kg) , and the recommended
test weight is 47 lb (23 kg). For the SB-150, the minimum test weight is 1.5 lb (0.7 kg) , and the
recommended test weight is 100 lb (50 kg).
Continued
SETUP MENU DESCRIPTIONS
"0" = Non-NTEP mode
"1" = NTEP mode
RS-232
REMOTE
PORT
DISPLAY
CONNECTOR
(SB-30 & 70 ONLY)
SB-30/70/150 Operation Manual
"1" = primary unit is in kg.
CALIBRATION MODE KEY FUNCTIONS
lb
ZERO
kg
GROSS
SET
CODE/VALUE
0
1√ √ √ √
100%√ √ √ √
1.9%
0√ √ √ √
1
Press ZERO key
to begin sequence
Press ZERO key
to begin sequence
Press the ZERO
key twice to exe-
cute.
NET
TARE
PRINT
9