APPENDIX C: Displayed Error Codes

CODE
Normal Operating Mode
_ _ _ _ _ _
Normal Operating Mode
Warmup Mode
– HI –
Warmup Mode
– LO –
Calibration Mode
Err 0
Err 1
Calibration Mode
Calibration Mode
Err 2
All Modes
Err 3
Err 4
All Modes
MODE
A weight greater than 102% of the scale's capacity has
been applied to the scale. Remove the weight from the
platter. Try re-calibrating the scale if this doesn't solve the
problem. Otherwise, possible load cell damage due to
overloading.
Underrange condition. Platter has been removed. Re-install
platter. Try re-calibrating the scale if this doesn't solve the
problem. Otherwise, possible load cell damage due to
shock loading.
The scale has been turned on with a weight already present
on the platter. Remove the weight and then press the ZERO
key if necessary. Try re-calibrating the scale if this doesn't
solve the problem. Otherwise, possible load cell damage
due to overloading.
The scale has been turned on without the platter installed.
Install the platter and then press the ZERO key if neces-
sary. Try re-calibrating the scale if this doesn't solve the
problem. Otherwise, possible load cell damage due to
shock loading.
Keyed-in weight value in Calibration Mode is less than 1%
of the scale's capacity. Use a larger test weight or re-adjust
value.
Keyed-in weight value in Calibration Mode is larger than the
scale's capacity. Use a smaller test weight or re-adjust
value.
Internal resolution is not high enough to process keyed-in
weight value in Calibration Mode. Verify test weight and
value.
Diagnostics check error - EEPROM Read
Diagnostics check error - EEPROM Write
SB-30/70/150 Operation Manual
MEANING / POSSIBLE SOLUTION
13