SECTION 13—CONTROL PANEL
11. Remove the On/Off switch. Refer to Replacing On/Off Switch on page 62 
12. Remove the circuit breaker by twisting the locknut on the outside of the control panel.
13. Discard existing control panel.
14. Install new control panel by reversing STEPS 3‐12.
15. Run unit and check for leaks at pressure sensor tubing on product tank cap. Refer to 
Leak Test on page 76
16. Reinstall cabinet. Refer to Removing Cabinet on page 23.
DETAIL "A"
Control
Panel
Tie
Wrap
Control Panel
Circuit Breaker
(STEP 9)
On/Off Switch
 
(STEP 9)
Perfecto
2
Tie
Wrap
P.C. Board
FIGURE 13.1 Removing Control Panel
54
DETAIL "B"
Control Panel
¼-inch Tubing (STEP 7)
Hour Meter
(STEP 9)
Mounting
Screws
Mounting Screws
(STEP 5)
¼-inch
Tubing
(STEP 6)
Part No 1154245