Invacare IRC5PO2AW Service Manual page 27

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

Top of Manifold
Exhaust Tube
Rubber
Grommet
6. Push the muffler assembly down through the opening in the sound box.
7. Cut the two tie wraps around the muffler assembly.
8. Separate the muffler assembly (Detail "A" in FIGURE 6.5).
9. Inspect the components to ensure they are clear.
10. Do one of the following:
• Replace exhaust canister and throttling muffler if plugged or restricted.
• Install only new throttling muffler if exhaust canister is in good condition.
11. Install new/existing muffler assembly by reversing STEPS 2‐10.
12. Reinstall cabinet. Refer to Removing Cabinet on page 23.
Part No 1154245
Bottom of Manifold
FIGURE 6.4 Replacing the Muffler Assembly
SECTION 6—PREVENTIVE MAINTENANCE
Exhaust Tube
27
Tie-Wrap
"F" Tube
Perfecto
2