Invacare IRC5PO2AW Service Manual page 40

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

SECTION 7—COMPRESSOR
Compressor
To avoid personal injury or property damage, rotate the fan by hand prior to connecting 
the unit to a power source. Check for suction at the air inlet by placing your finger over 
the port as you rotate the fan. You should feel a slight suction with each rotation of the 
fan. If you DO NOT feel a suction but you feel or hear a thump as you turn the fan, DO 
NOT connect the unit to a power source. Review the assembly procedure for possible 
error.
NOTE: For this procedure, refer to FIGURE 7.7 on page 41.
1. Reinstall compressor. Refer to Replacing Compressor Assembly on page 31. 
2. Run unit and check for leaks around compressor hoses and compressor tubes. Refer 
to Leak Test on page 76.
3. Take pressure readings to ensure proper system pressures. Refer to Checking Sieve 
Bed Pressure on page 85.
Perfecto
2
WARNING
40
Part No 1154245