Invacare IRC5PO2AW Service Manual page 9

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

Placement:
Tubing:
Relative Humidity:
Time of Operation:
Recommended Storage and Shipping Temper-
ature:
Standards and Regulatory Listing:
CE marked model
LABEL
SYMBOL
O
LABEL
SYMBOL
Part No 1154245
No closer than 7.62 cm (3 inches) from any wall, furniture, drap-
eries, or similar surfaces
2.1 m (7 ft) cannula with a maximum 15.2 m (50 ft) of
Crush-Proof Tubing (DO NOT pinch)
20 to 60%
Up to 24 hours per day
-29°C to 65°C (-20°F to 150°F), at 15-95% relative humidity
IRC5PO2AW/IRC5PAW
IEC/EN61000-3-2
IEC/EN61000-3-3
IEC/EN 60601-1, A1, A2
IEC/EN 60601-1-2
IRC5PO2AW Models Only
ISO8359
MDD
IRC5PO2AW
IRC5PO2AW INDICATORS
O
PURITY
2
SYSTEM OKAY
O
over 85% (± 2%)
2
2
O
Between 73% (± 3%)
2
to 85% (±2%)
SYSTEM FAILURE
O
Below 73% (±3%)
2
IRC5PAW INDICATORS
STATUS
SYSTEM OKAY
SYSTEM FAILURE
TYPICAL PRODUCT PARAMETERS
INDICATOR LIGHTS
(LED)
GREEN Indicator Light
YELLOW Indicator light
A. YELLOW Solid
B. YELLOW Flashing Sensor
Failure
Call a qualified technician.
RED Indicator Light
Continuous Audible Alarm
Sieve-GARD Compressor Shut-
down.
INDICATOR LIGHTS
(LED)
GREEN Indicator Light
RED Indicator light
Continuous Audible Alarm
Sieve-GARD™Compressor
Shutdown
Call a qualified technician
9