Invacare IRC5PO2AW Service Manual page 19

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

SYMPTOM
Unit not operating: Alarm:
On or Off
Internal Status Indicators:
RED: Two Flashes
GREEN: Four Flashes Or
RED: Two Flashes GREEN:
Five Flashes
Low Concentration: NOTE:
Check for O
purity using a
2
calibrated Oxygen Analyzer
at Test Point 1 (oxygen
outlet) of the concentrator.
Internal Status Indicators:
RED: Two Flashes
GREEN: One Flash73%
Shutdown Control Panel
Indicators: SensO
ONLY:
2
RED: On
YELLOW: Off
GREEN: Off
For SensO
units, the RED
2
indicator will signal extremely
low purity and will be
accompanied by a continuous
audible alarm and a system
shutdown. Repairs are
required.
Part No 1154245
PROBABLE CAUSE
P.C. board Failure.
Cabinet filters dirty.
Compressor inlet filter dirty.
Compressor:
a. Defective.
b. Faulty capacitor.
c. Bad motor windings.
d. Worn seals.
e. Bad bearings.
f. Leak at fittings or tubing.
g. Leaky or defective relief valve.
h. Insufficient voltage (outlet).
Heat exchanger:
a. Leak at tubing or body chamber.
b. Inspect tubing and heat
exchanger.
Regulator cracked or leaking.
Exhaust muffler dirty or plugged.
Fan:
a. Not operating. Unit overheating.
b. Faulty fan.
Sieve beds defective.
Tubing kinked or blocked.
P.C. board:
a. Shifts at wrong pressures.
19
SECTION 4—TROUBLESHOOTING
SOLUTION
Replace P.C. board.Refer to Replacing P.C.
Board on page 58.
Clean or replace. Refer to Cleaning the
Cabinet Filter on page 24, Replacing the
Outlet HEPA Filter on page 25, and
Replacing the Compressor Inlet HEPA
Filter on page 26.
Replace inlet filter. Refer to Replacing the
Compressor Inlet HEPA Filter on page 26.
a. Replace compressor. Refer to
Replacing Compressor Assembly on
page 31.
b. Replace capacitor. Refer to Replacing
Capacitor on page 33.
c. Replace compressor.
d. Rebuild top end of compressor. Refer
to Rebuilding the Thomas Model
2660 Compressor on page 34.
e. Replace compressor.
f. Replace fittings or tubing.
g. Replace relief valve.
h. DO NOT use extension cords.
a. Replace or retighten.
b. Replace or retighten tubing. Replace
heat exchanger. Refer to Replacing
Heat Exchanger Assembly on
page 51.
Replace. Refer to Replacing Regulator on
page 48.
Replace. Refer to Replacing the Muffler
Assembly on page 26.
a. Leads to fan disconnected.
Reconnect.
b. Replace.Refer to Replacing Cooling
Fan on page 55.
Replace. Refer to Checking Sieve Bed
Pressure on page 85.
Repair or replace.
Check pressure at product tank. Pressure
should rise to 144.79 kPa (21 p.s.i.) at shift
point. If not, replace P.C. Board. Refer to
Replacing P.C. Board on page 58.
Perfecto
2