Removing Compressor and Compressor Heads

Compressor
1. Unplug concentrator and carefully remove compressor from concentrator. Refer to 
Replacing Compressor Assembly on page 31.
Compressor Head Plate
NOTE: For this procedure, refer to FIGURE 7.3.
1. Clean loose dirt from the outside of the 
compressor. 
2. Use the Torx T‐25 torque wrench and 
loosen the eight head screws.
3. Note head placement before removing 
compressor head plate.
4. Remove the compressor head plate.
Valve Plates
NOTE: For this procedure, refer to 
FIGURE 7.3.
1. Note the position of the plates before 
removing.
2. Lift off the valve plate on one side.
3. Repeat for the other side.
NOTE: Note the position of the tab on valve 
plate for reassembly.
NOTE: To avoid confusion, ONLY service one 
end of the compressor at a time.
Part No 1154245
SECTION 7—COMPRESSOR
Compressor
FIGURE 7.3 Removing Compressor and
Compressor Heads - Rebuilding the
Compressor - Compressor - Compressor
Head Plate - Valve Plates
35
Head Screws
Compressor
Head
Plate
Valve
Plate
Perfecto
2