Invacare IRC5PO2AW Service Manual page 36

Oxygen concentrator homefill compatible
Hide thumbs Also See for IRC5PO2AW:
Table of Contents

Advertisement

SECTION 7—COMPRESSOR
Fan
NOTE: For this procedure, refer to FIGURE 7.4.
1. Using two flat blade screwdrivers, pry the fan off of the motor shaft under the center 
housing. DO NOT pry under the fan blades.
NOTE: Note position/orientation of fan for reassembly.
Connecting Rod and Eccentric Assembly
NOTE: For this procedure, refer to FIGURE 7.4.
1. Rotate shaft to align the eccentricʹs set screw with the access hole in the bottom of the 
compressor housing.
2. Insert the 1/8‐inch Allen wrench into the access hole in the bottom of the compressor 
housing.
3. Loosen the set screw ¼ turn.
4. Rotate connecting rod to top dead center (180°) and slide the connecting rod/eccentric 
assembly off the shaft through the opening in the housing.
5. Secure the rod assembly in a fixture.
6. Remove the sleeve from the connecting rod and discard.
Piston Cup
NOTE: For this procedure, refer to FIGURE 7.4.
1. Remove the torque screw (discard) from the piston cup retainer (retain for reassembly).
2. Remove the piston cup (discard) and wipe debris from the top of the connecting rod 
with a clean damp cloth.
NOTE: Partial 
compressor shown.
Eccentric Set
Screw Access Hole
FIGURE 7.4 Removing Compressor and Compressor Heads - Rebuilding the Compressor -
Fan - Connecting Rod and Eccentric Assembly - Piston Cup
Perfecto
2
Cylinder Sleeve
Torque Screw
Retainer
Piston Cup
Rod Top
Connecting Rod
Eccentric
Fan
Set Screw (hidden)
36
NOTE: As the piston travels up and 
down it will also rock from side‐to‐side. 
This is a feature of the WOB‐L Piston.
Part No 1154245