On-Screen Messages; Messages Explained - Lifeloc FC10Plus Operation Manual

Professional breath alcohol tester
Table of Contents

Advertisement

Lifeloc FC10

Messages Explained

Message
<1.3L Retest or Try Manual Test
>0.6
Calibration/Cal Check Expired
Calibration/Cal Check Expiring
External Interference
Flow Error - Retry & Blow Steadily
Invalid Calibration/Cal Check
Log Empty
Low Battery
Low Li Battery
Pump Reset Needed
Temperature
Timeout
Plus
www.Lifeloc.com

On-screen Messages

Explanation
Breath fl ow ended before the subject blew
1.3 liters of breath. Instruct subject to try
again or use manual test mode.
BAC is unusually high and above 0.6 BAC.
Subject may require medical attention.
Calibrate or Cal Check the unit.
The unit is within 48 hours of the
Calibration or Cal Check lockout time.
External interference has been detected.
Move to a different location and try again.
Exhalation not complete or interrupted.
Instruct subject to blow steadily as long as
they can.
No alcohol was detected. Repeat setup,
then retest.
There are no results in the memory.
Battery voltage is too low to take a test.
Replace batteries.
The internal clock battery is low. Contact
Lifeloc Technical Support.
The pump needs to be reset. Follow
onscreen instructions.
When calibrating or taking a test, unit is
outside of temperature limits.
Calibration range = 68
o
Testing range = 32
User blew for > 15 seconds.
Try again, blow harder.
|
303-431-9500
28
o
o
o
- 95
F (20
- 35
o
o
o
- 130
F (0
- 55
C)
o
C)