Wet Bath Calibration; Wet Bath Simulator Set-Up; Wet Calibration - Lifeloc FC10Plus Operation Manual

Professional breath alcohol tester
Table of Contents

Advertisement

Lifeloc FC10

Wet Bath Simulator Set-Up

Pour a bottle of certifi ed alcohol solution into the simulator jar and hand tighten lid.
Connect long tube from input port to output port so no alcohol escapes while simulator
heats up.
Plug in the simulator and turn it on.
The simulator automatically heats the solution to 34° C (93.2° F) in about 5 to 10 minutes.
Proper operating temperature is important for accuracy so be certain to check
temperature before proceeding.
Performing a Wet Bath Calibration
Prepare the wet bath simulator according to
its instructions above.
Disconnect long tube from the output port.
Attach a mouthpiece adapter securely to the
output port on the simulator lid. Refer to your
simulator manual for location.
Turn the FC10
Plus
on. Attach a mouthpiece to
the back of the unit.
Press the Function button until the display
reads "CALIBRATION", then press the
Execute button. Display reads "WET CHECK".
Press the Function button until the display
reads "WET CALIBRATE".
Verify the "CAL STANDARD" is set to the concentration of certifi ed alcohol solution you
will be using when you calibrate. To select a new Calibration Standard, see page 24.
Plus
Slide the FC10
mouthpiece over the mouthpiece adaptor on the simulator.
Be prepared to blow into the tube for up to 10 seconds.
Start blowing
• Blow through the input tube (or use a calibration pump) to create and maintain 1/2" of
bubbles on the surface of the solution.
• Press Execute to proceed and start a 3 second countdown.
• Press Execute to take a sample.
• Continue blowing for another 3 seconds.
Stop blowing
Plus
For Calibration & Calibration Check
www.Lifeloc.com

Wet Bath Calibration

Calibration
Pump
|
303-431-9500
22
Simulator