Lifeloc FC10
Selecting the Calibration Standard
Press the Function button until the display reads "CALIBRATION".
Press the Execute button. The display reads either "WET CHECK" or " DRY CHECK".
Press the Function button until the display reads "CAL SETTINGS".
Press the Execute button. The display reads "CAL STANDARD".
Use the + or – button to change the number to the BAC level of standard you will be
calibrating to. It should be the same as on the bottle of certifi ed solution or as your
altitude corrected standard, if using a dry gas tank. (See p. 18)
Press the Function button to save the settings. Momentarily press Power to return to the
testing mode.
Once you set the standard, you do not have to set it again unless you change solutions or
elevation (dry gas only).
Calibration/Cal Check Explained
Calibration of an FC10
accurate BAC (Breath Alcohol Concentration) results.
You can use dry gas or wet bath solution to calibrate your FC10
solutions are .100, .080 or .040 BAC.
Plus
The FC10
must be between 68° and 95° F (20° - 35° C) to calibrate.
Lifeloc recommends you calibrate your FC10
Once every 12 months, regardless of how many tests you have performed
Or, at intervals specifi ed by your Internal Policies, Quality Assurance Plan, or State
Regulations
Or, after two failed Calibration Checks.
Plus
Plus
sets your unit to a known alcohol concentration to enable
www.Lifeloc.com
Calibration Settings
Plus
Plus
:
|
303-431-9500
19
. Most commonly used