A Calibration Check simply verifi es the FC10
acceptable accuracy range. Calibration check is also referred to as "External Calibration
Check," "Accuracy Check," "Verifi cation" and "Cal Check."
Lifeloc recommends you perform a calibration check on your FC10
Once every 30 days
Or, at intervals specifi ed by your Internal Policies, Quality Assurance Plan, or
State Regulations.

Selecting the Standard Type

Press the Function button until the display
reads "CALIBRATION".
Press the Execute button. The display reads
either "WET CHECK" or "DRY CHECK."
Press the Function button until the display
reads "CAL SETTINGS".
Press the Execute button. The display reads
"CAL STANDARD".
Press the Function button until the display
reads "STANDARD TYPE".
Press the Execute button to choose between
"DRY GAS" and "WET BATH".
Press the Function button to save settings. Momentarily press Power to return to the
testing mode.

Calibration/Cal Check Lock-out Explained

The calibration/cal check lock-out allows you to prevent usage of your FC10
calibrated or cal checked in a specifi ed time period. Time can be registered as days or
number of tests.
Beginning 48 hours before the specifi ed lock-out time, the unit will display "WARNING
CAL EXPIRING," "WARNING CHECK EXPIRING" or both.
When the time period has elapsed, the unit will display "CAL EXPIRED," "CHECK
EXPIRED" or both and not allow testing until it is calibrated and/or cal checked.
You can specify time periods up to 999 days or 9999 tests between calibrations or cal
checks. You can also disable the calibration/cal check lock-out feature. (See p.21)