Fuel Cells; Cleaning; Batteries - Lifeloc FC10Plus Operation Manual

Professional breath alcohol tester
Table of Contents

Advertisement

Lifeloc FC10

Fuel Cells

Fuel cells are highly durable sensors that are capable of providing accurate breath alcohol
results for years. There are, however, a few precautions you should take to make certain
that these devices perform for the longest period of time possible.
Use the device. Fuel cells like moisture, so it is a good idea to take tests periodically
to provide needed moisture to the fuel cell, especially in dry climates. You do not need
alcohol, just breath.
Avoid cigarette smoke! Make certain no one is permitted to blow cigarette smoke into
the unit. This can damage or destroy the fuel cell.

Cleaning

Use of a mild disinfectant cleaner and a soft cloth on the outside of the case is
recommended periodically to keep your unit clean. Do not use alcohol to clean the unit.
Lifeloc offers alcohol free cleaning wipes specifi c for your unit. See page 31 for more
details.

Batteries

Your FC10
Plus
default setting is for four AA batteries. However, you do have the ability to
use NiMH rechargeable batteries as well.
To use NiMH, insert the batteries as directed and power on the unit.
Press the Function button until the display reads "SETTINGS".
Press the Execute button. Display shows "TIME".
Press the Function button until the display shows "BATTERY TYPE".
Press the Execute button to change your battery settings.
Press the Function button to save your settings.
Momentarily press the Power button to return to the testing mode.
The four AA batteries in your FC10
6000 tests. It is recommended you use high-quality alkaline batteries with your unit.
Plus
Plus
should last for about 160 hours of "on" time or up to
www.Lifeloc.com
|