Setting The Calibration Lock-Out Type; Setting The Cal Check Lock-Out Type - Lifeloc FC10Plus Operation Manual

Professional breath alcohol tester
Table of Contents

Advertisement

Lifeloc FC10

Setting the Calibration Lock-out Type

Press the Function button until the display reads "CALIBRATION".
Press the Execute button. Display reads either "WET CHECK" or "DRY CHECK".
Press the Function button until the display reads "CAL SETTINGS".
Press the Execute button. Display reads "CAL STANDARD".
Press the Function button until the display reads "CAL TIME TYPE".
Press the Execute button to select "DAYS" or "TESTS".
Press the Function button to save your setting.
Display will now read "CAL TIME", or "CAL NUM TESTS".
Press the + or - button to set the number of days or tests between calibrations. The time
starts counting from your last calibration date, not from the date you set it. Choosing
"DISABLED" turns this setting off.
Press the Function button to save your setting.
Momentarily press Power to return to the testing mode.

Setting the Cal Check Lock-out Type

Press the Function button until the display reads "CALIBRATION".
Press the Execute button. Display reads either "WET CHECK" or "DRY CHECK".
Press the Function button until the display reads "CAL SETTINGS".
Press the Execute button. Display reads "CAL STANDARD".
Press the Function button until the display reads "CHECK TIME", or "CHECK NUM TESTS".
Press the + or - button to set the number of days between calibration checks. The days
start counting from your last calibration check date, not from the date you set it.
Press the Function button to save your settings.
Momentarily press Power to return to the testing mode.
Plus
www.Lifeloc.com
Calibration Settings
|
303-431-9500
21