10.0 CALIBRATION

ACTION
Press the [Tare/Zero] key during the self-
test at power on. The scale will ask to
enter the password.
Default Password is 0000.
Enter "0" four times.
Password can be changed in technical
parameters.
Press [Print].
Select the scale to be set up by using the
[Local/Remote] key.
Press the [Print] key to enter the
technical section.
Use the [U.Wt.] to select the weighing unit
to be used for setting the calibration for
the scale. The arrow in the "Weight"
window will indicate the unit selected.
Press the [Print] key to continue.
You will enter the first parameter -
Calibration. Display will ask to unload any
weight on the platform.
Press the [Print] key to continue.
Load the calibration weight onto the scale
If the mass is different than the value
shown press the [CE] key then enter the
calibration weight to be used .
Then press [Print].
© Adam Equipment Company 2011
DISPLAYS
" Pi n"
" Pi n" " - - - -"
"tECH" " LocAL" "
"tECH" "rEmo tE" "
"tECH" " Uni t" "
"nO L o Ad"
"LoAd " " 0010" " KiLoS"