Invacare Bora Service Manual page 19

Hide thumbs Also See for Bora:
Table of Contents

Advertisement

The figure shows which LEDs are located on the
remote.
1) Battery display
2) Profile indicator
3) Adjusting motors
Error code
1 LED battery display
2 LED battery display
3 LED battery display
4 LED battery display
5 LED battery display
6 LED battery display
7 LED battery display
8 LED battery display
9 LED battery display
10 LED battery display
7 LED Battery display
plus
5 LED
profile indicator
POSSIBLE
SOLUTION
CAUSE
Batteries
Charge battery
discharged
Check cable to batteries
Error at left-hand
Check connection plug, check
motor
motor
Connection
loose/defective
or motor
defective
Short-circuit in
Check connection plug, check
left-hand motor
motor
Error at right-
Check connection plug, check
hand motor
motor
Connection
loose/defective
or motor
defective
Short-circuit in
Check connection plug, check
right-hand motor
motor
The mobility
Remove battery charger
device has been
blocked by an
external signal,
for example
because the
charger is
connected.
Fault on drive
Put the drive lever in a central
lever
position before switching the
remote on.
Fault in the
Check cable
electronics
The parking
Check parking brake
brake is not
Check cable.
working correctly.
Power surges in
Check cable to batteries
the control box,
e.g. due to bad
connection with
batteries.
Compatibility
Check cable to remote
problems
Replace remote
between
modules
Documentation
See chapter 7.7.
See chapters 7.7
and 7.2
See chapters 7.7
and 7.2
See chapters 7.7
and 7.2
See chapters 7.7
and 7.2
See chapter 7.7.
See chapter 7.7.
See chapter 7.7.
See chapter 7.7.

Hide quick links: