10. Start-up

The system must have warmed up, or cooled down, to room tempera-
ture before start-up can be begun.
Check that all hose connections have been correctly made as described
in the "Setting up" section.
Switch the system on by plugging the mains plug in a suitable electric socket.
After a brief self test the system begins the production of ultrapure water.
Wait until the first tank filling has been completed and discard the water that is in
the tank. Subsequently check the concentrate flow and adjust this as described in
the "How Smart2Pure functions" section.
Withdraw at least 1 litre of water before you screw the sterile filter (10) that is
standardly supplied in the outlet of the dispensing valve.
Now set the wanted display mode, the limiting value for the conductivity as well as
the limiting value for the temperature as described in the "System control" section.