Error Messages - Leica IP S Instructions For Use Manual

Table of Contents

Advertisement

7.3 Error messages

Display
code
Magazine output mechanically blocked.
Faulty ejection of a slide. Magazine output blocked.
Transfer of a slide from the chute to the specimen
slide carrier failed. Horizontal motor either incorrectly
positioned or mechanically blocked.
Slide is stuck in the feeding chute.
Horizontal drive is mechanically blocked.
Vertical drive is mechanically blocked.
Rotational motion is mechanically blocked.
Slide not correctly clamped in specimen slide carrier.
Slide did leave the feeding chute but did not reach the
specimen slide carrier.
Slide did not leave the specimen slide carrier or was
still located in the specimen slide carrier during initial-
ization.
The ink print head is getting too hot.
Ambient temperature too high or electronics defective.
No or incorrect voltage at ink print head.
Leica IP S
Source of error
7.
Troubleshooting
Troubleshooting solution
Remove the cause of the blockage.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide from the specimen
slide carrier.
Remove the slide from the specimen
slide carrier.
Switch instrument off and let it cool off.
The instrument remains disabled until
the print head has cooled down to a
temperature value within the permis-
sible range.
Check ambient temperature.
Call Leica Technical Service.
61