Technical Data - Leica IP S Instructions For Use Manual

Table of Contents

Advertisement

3.
Instrument Components and Specifications

3.2 Technical data

General
Approvals:
Nominal supply voltages:
Nominal frequency:
Power fuse:
Maximum power draw at 100 - 120 V:
Maximum power draw at 200 - 240 V:
Leakage current at 240 V/50 Hz:
Nominal power:
IEC 1010 classification:
Operating elevation:
A-weighted noise level:
IP protection class (IEC 60529)
Climatic conditions for the operation of the instrument:
Temperature:
Relative humidity:
Climatic conditions for the storage and transport of the packaged instrument:
Storage temperature range:
Transport temperature range:
Relative humidity:
Dimensions and weight
Dimensions of basic instrument
Width x depth:
Height with magazine:
Dimensions with unload station connected
Width x depth:
Height with magazine:
Basic instrument empty weight:
Weight, packed:
Unload station empty weight:
Weight, packed:
14
The instrument-specific approval marks are located at the rear panel
of the instrument, next to the name plate.
100 to 120 V ~ +/- 10 %
200 to 240 V ~ +/- 10 %
50 to 60 Hz
2x T 3.15 A L250 V
4.0 A
2.8 A
ca. 2.4 mA
700 VA
Protection class 1, pollution degree 2
Overvoltage installation category II
up to max. 2000 m NN
< 70 dB (A)
IP20
+15 °C to +30 °C
20 - 80 % - non-condensing
+5 °C to +40 °C
-29 °C to +40 °C
10 - 85 % - non-condensing
475 x 650 mm
560 mm
548 x 650 mm
655 mm
approx. 28 kg
approx. 65 kg
approx. 14 kg
approx. 32 kg
Instructions for Use V 1.9 RevD - 05/2013