Error Messages - Leica IP S Operating Instructions Manual

Automated printing system for microscope slides
Table of Contents

Advertisement

7.
Troubleshooting

7.3 Error messages

Display
code
20
Magazine output mechanically blocked.
21
Faulty ejection of a slide.
Magazine output blocked.
22
Transfer of a slide from the chute to the slide
carrier failed. Horizontal motor either
incorrectly positioned or mechanically blocked.
23
Slide is stuck in the feeding chute.
32
Horizontal drive is mechanically blocked.
33
Vertical drive is mechanically blocked.
34
Rotational motion is mechanically blocked.
35
Slide not correctly clamped in slide carrier.
Slide did leave the feeding chute but did not
reach the slide carrier.
36
Slide did not leave the slide carrier or was still located
in the slide carrier during initialization.
37
The ink print head is getting too hot.
Ambient temperature too high or electronics defective.
38
No or incorrect voltage at ink print head.
50
Source of error
Troubleshooting solution
Remove the cause of the blockage
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide.
Remove the slide from the slide carrier.
Remove the slide from the slide carrier.
Switch instrument off and let it cool off.
The instrument remains disabled until
the print head has cooled down to a
temperature value within the permissi-
ble range.
Check ambient temperature.
Call Leica Technical Service.
Instruction manual V 1.6 – 07/2005

Advertisement

Table of Contents
loading