Error messages
(continued)
Display
code
Faulty control data received (program bug)
Settings for the serial interface incorrect or the instru-
ment configuration conflicts with the PC configuration.
Transmitted data do not contain confirmation of
receipt or data transmission was not confirmed by
the PC.
Print image exceeding the vertical limit.
Print image exceeding the horizontal limit.
The CRC test of the EEPROM returned an error when
the instrument was switched on.
Internal firmware error or
...
defective controller.
Mechanical problem makes ejection of cassettes from
...
magazine difficult.
A print job was attempted with a storage fluid cartridge
inserted.
Power supply unit does not achieve standard operat-
ing voltage.
Firmware only partially loaded or not loaded at all.
Flash memory defective.
Wrong firmware.
Leica IP S
Source of error
7.
Troubleshooting
Troubleshooting solution
Carry out a RESET on the printer.
Check cable connection to the PC.
Check configuration of serial port of
the PC and reboot the PC.
Follow the same procedure as for
ERROR 60.
Try another printer cable.
Error caused by application software.
Error caused by application software.
Call Leica Technical Service.
Call Leica Technical Service.
Check ejection mechanism. Remove
foreign bodies, then clean with brush.
Remove storage cartridge and insert
ink cartridge. Insert ink cartridge and
press LOADED to confirm. (See
ter
4.9)
Call Leica Technical Service.
Call Leica Technical Service.
Call Leica Technical Service.
Chap-
63