Honeywell Sensepoint XCD Technical Manual page 69

Gas detector
Hide thumbs Also See for Sensepoint XCD:
Table of Contents

Advertisement

Sensepoint XCD Technical Manual
Notes:
1. Since catalytic sensors require oxygen for correct operation, a mixture of gas in air
should always be used for calibration purposes.
2. Assuming average performance of the sensor, the sensitivity information in Tables 11 To
13 is normally accurate to + or - 20%.
Working Example:
If the target gas to be detected is 0-100%LEL Ethylene and the only calibration gas
available to re-calibrate the sensor is Methane (at 50% LEL concentration), the procedure
is as follows:
(1) Look up the star rating for each gas in Table 11:
Gas No. 14, Ethylene = 5*
Gas no. 18, Methane = 6*
(2) Then, look up the span settings for a 50% LEL calibration gas in Table 12 by selecting
the row of figures next to the 6* in the "calibration gas" column. Select the figure in the 5*
column of the "gas to be detected" section. The figure is 62.
(3) This means that during re-calibration, the span gas setting on the gas detector
transmitter or controller should be set to 62% LEL to give an accurate measuring scale for
0-100%LEL Ethylene, when using 50% LEL Methane as the calibration gas.
Please contact your local Honeywell Analytics sales or service distributor, or regional office
should further clarification or additional information be required.
69
SPXCDHMANEN Issue 2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents