Honeywell Sensepoint XCD Technical Manual page 36

Gas detector
Hide thumbs Also See for Sensepoint XCD:
Table of Contents

Advertisement

Sensepoint XCD Technical Manual
Menu
Set Calibration
Select Sensor
Select Gas
Set Range
Configure Inhibit
Set Password
Set Calibration Interval
Bump Test
Force Current
Set Alarms
Set Relays
Display
Description
Execute zero/span calibration
Set calibration gas level
After zero, the option exists to proceed with
span calibration, or return to the Menu.
Select the type of sensor from the sensor
list. This menu is only available for
flammable / IR sensor.
Select the type of gas from the list. This
menu is only available for flammable / IR
sensor.
Set measuring range
Select inhibit current
Set timeout option (5 minute increments)
Enable/disable password
Set password
Default – no password (Select '0000')
Set calibration interval, 30 to 365 days
User configurable option to display warning
Execute a 'bump' test to check gas
response of the sensor.
Force analogue output to test functionality
of GD control system during system
commissioning.
Set alarm 1, alarm 2 levels, functionality
and operation (none/falling/rising)
Set relay 1,2,3 type (alarm 1, alarm 2,
fault and inhibit) and action (energized/de-
energized)
36
SPXCDHMANEN Issue 2

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents