10.2

Adjusting forks

1. Remove the headtube cover from the castor headtube.
2. To properly tighten castor journal system and guard against flutter, perform the following check:
Tip back the wheelchair to floor.
Pivot both forks and castors to top of their arc simultaneously.
Let castors drop to bottom of arc (wheels should swing once to one-side, then immediately
rest in a straight downward position).
Adjust locknuts according to freedom of
castor swing.
3. Test wheelchair for manoeuvrability.
4. Readjust locknuts if necessary, and repeat
STEPS 1-3 until correct.
5. Snap headtube cover into the castor
headtube.
76
Fork
NOTE: Components exploded for clarity. There is no need
to remove the fork from the base frame.